Particle Size Analyzer LA-960

Sebuah produk Mesin untuk untuk mengukur ukuran partikel yang ideal untuk digunakan di laboratorium adalah Particle Size Analyzer LA-960. Instrument ini bisa diandalkan dalam proses mengukur ukuran partikel, zeta potensial dan berat molekul. Bagi anda yang membutuhkan diskusi cepat produk ini bisa menghubungi contact PT. Andaru Analitika Sains via whatsapp 087777277740 atau telepon 0251-7504679.

andaru analitika sains produk particle size analyzer la 960

Berikut adalah informasi produk Particle Size LA-960  yang bisa anda pelajari :

Overview Product Particle Size Analyzer LA-960

HORIBA proudly introduces the newest breakthrough in particle size technology.

This latest evolution in the LA series advances scientific knowledge for tomorrow’s world through intuitive software, unique accessories, and high performance.

The LA-960V2 continues HORIBA’s proud tradition of leading the industry with innovative design.

HORIBA’s groundbreaking optical design perfects the static light scattering particle sizing technique.

  • Advanced Detector Design
  • Superior Instrument-to-Instrument Precision
  • Automatic Laser Alignment in Seconds
  • Guaranteed Accuracy and Precision

State of the Art Sampling Systems

Wet Measurement

Sample-to-sample analysis in less than 60 seconds

The HORIBA LA-960V2 wet circulation system is an easy, fast and very powerful dispersion system. The standard wet system offers a full package of a dispersant fill pump, liquid level sensor, circulation pump, 30 W in-line ultrasonic probe, and drain valve, which is all software-controlled for true one-button operation. This advanced design provides highly reproducible particle size results.

Dry Measurement

Automated, powerful dry-powder dispersion

The LA-960V2 PowderJet combines several unique and patented features to provide the most reproducible dry measurements. Use the Auto Measurement function to control vacuum, air pressure, powder flow, start/stop conditions, measurement duration, and data processing. Designed to handle every application including small sample amounts, friable powders, and highly agglomerated materials.

  • Smart Scans – Trigger function
    This function allows for very precious materials to be measured accurately. When the sample amount is limited or low flow ability, the Trigger functions perfectly to start and stop the measurement.
  • Self-Adjusting Powder Flow
    Historically, the biggest challenge in dry powder measurement involved establishing an even powder flow. The LA-960V2 PowderJet has solved that challenge with a self-adjusting feedback loop to maintain a constant laser transmittance. This is a crucial factor in creating reliable, reproducible dry powder size measurements.

Diagram LY-9610 Imaging Unit

The imaging unit allows real-time observation, particle image acquisition, and assessment of the particles in the wet circulation system. It is small and integrated in the main unit without increasing instrument footprint!

diagram ly 9610

Features Particle Size Analyzer LA-960

  • Detect and count even with trace amounts of unusual particles (Contaminants, large particles and aggregated particles)
  • Detect bubbles in wet circulation system.
  • Improved understanding of sample dispersion (effect of stirring, sonication, addition of surfactants, stabilizers, etc.)
  • Particle shape analysis* (Optional Advanced Software)

Functions Particle Size LA-960

  • Image acquisition and particle detection.
  • Automatic detection of large particles.
  • Automatic detection of bubbles.
  • Real time trend graph and histogram of equivalent circular diameter.
  • Histogram of shape distribution (aspect ratio, roundness, length, width) with optional advanced software.

Specifications

Measurement Principle Mie scattering and Fraunhofer diffraction
Measurement range 10 nm – 5000 μm
Measurement Time Typical measurement takes 60 seconds from liquid filling, sampling and measurement to rinsing.
Measurement Method Circulation measurement or fraction cell measurement (Fraction cell is optional)
Sample Quantity Approximately 10 mg – 5 g (Depending on the particle size, distribution and density)
Dispersing Volume Approximately 180 mL for standard pumping system, 5/10/15 mL for FractionCell accessory, Manual filling : 35 mL, Automatic filling : 40 mL for MiniFlow accessory. Approximately 1 L of LiterFlow option.
Available carrier fluid Aqua* (A type), Organic solvent (S type) (*Ethanol can be used as a dispersing additive)
Communication USB 2.0
Light Sources Red solid state 5 mW laser diode (650 nm), Blue solid state 3 mW LED (405 nm)
Dispersion System In-line ultrasonic probe: 30 W, 20 kHz, adjustable levelsCirculation pump: Fully automated fill and circulation pumps, 15 adjustable speeds, 4 selectable fill levels, 15 selectable circulation speeds (max: 10 L/min)
Operating Conditions 15〜35°C (59 to 95°F), relative humidity 85% or less (no condensation)
Power AC 100-240V 50/60Hz, 300VA
Dimensions 705 (W) × 565 (D) × 500 (H) mm
Mass 54 kg
Computer Requirements PC operation, Software compatible with Windows® 10 32-bit and 64-bit environments,

Demikian informasi produk Particle Size Analyzer LA-9610. Bagi anda yang membutuhkan diskusi cepat produk ini bisa menghubungi contact PT. Andaru Analitika Sains via whatsapp 087777277740 atau telepon 0251-7504679. Link alamat kami sertakan pada gmaps.